Forster Resonance Energy Transfer (FRET)

An exciting application of fluorescence

Solutions

FRET, or Forster Resonance Energy Transfer, is a phenomenon where closely matched pairs of fluorophores are used to determine spatial proximity in molecular and protein-protein interactions. Three criteria must be met for efficient FRET to occur between the donor and acceptor chromophores:

  1. Strong spectral overlap between the emission spectrum of the donor and the excitation spectrum of the acceptor
  2. Close proximity of the donor and acceptor chromophores (<10 nm)
  3. Alignment of the transition dipoles of the two molecules

When these criteria are met, the energy is non-radiatively transferred from the donor to the acceptor. As a result, the donor’s fluorescence intensity goes down (and it’s excited-state lifetime is shortened) and the acceptor’s fluorescence intensity goes up.

As described above, the donor and acceptor chromophores need to have a large spectral overlap, which, while creating the conditions for effective energy transfer, also results in spectral bleed through (SBT) and can be a problem in FRET measurements. Because excitation and emission spectra are very wide, there is bleed through in both directions. A small percentage of acceptor molecules can be directly excited by the donor’s excitation, and simultaneously, a small percentage of donor emission can pass through the acceptor’s emission filter.

The development of SBT correction techniques have been critical to the evolution of FRET as a useful and widely used application. These SBT correction techniques—which include software development, fluorescence lifetime imaging (FLIM) correlation, and photo bleaching techniques—have reached a degree of sophistication that improves the efficacy of FRET. Similarly, the development of microscopy techniques such as one-photon, two-photon (multiphoton), confocal, and TIRF are all contributing to the growing effectiveness and ease of FRET experiments.

Fluorescent microscope

Filter & Microscope Configurations for FRET

The filter components required for FRET experiments are not esoteric. As in any fluorescence microscopy application, an excitation filter is required for exciting the donor fluorophore, and a dichroic mirror is required for separating donor excitation energy from both donor and acceptor emission energy. Unlike typical fluorescence applications, however, two emission filters are required, one for the acceptor fluorophore, or FRET emission, and one for the donor fluorophore in order to correct for signal bleed thru. For the most part, the filter components used for FRET are identical to those used for epifluorescence applications, providing that the dichroic mirror transmits both emission wavelengths. More important than the selection of filters is an understanding of the physical configuration of the microscope hardware to be used in the FRET experiment. At issue are critical experimental variables, such as time and image registration. Grating While the ideal setup may not be affordable or available to all researchers interested in FRET studies, it is nonetheless important to understand the pros and cons of the available hardware and filter set configurations.

Plate BeamsplitterFor dynamic imaging of FRET due to protein-protein or other intermolecular interactions, the preferred setup allows the simultaneous visualization of both donor and acceptor signals. This arrangement is achieved with a series of dichroic mirrors and detectors arranged in a sequential manner such that different wavelengths are imaged onto each detector at the same time. The detectors are aligned so they are in perfect registration. Real-time corrections of bleed through can be accomplished using this method. Because the signals are acquired at the same time, it is the best method to use for imaging of live specimens. It is costly to set up, but robust because there are no moving parts in the system.

The next best option is to use a filter wheel for fast-switching between donor and acceptor emission filters. When using a filter wheel in front of the detector, the filter cube is equipped with an excitation filter and dichroic mirror with the emission side left empty. When the donor and acceptor filters are loaded in the filter wheel, switching can be accomplished in ~50 ms which is fast enough for many biological processes. Because a single detector is used, this method is less expensive, but there may be subtle shifts in image registration due to movement of the filter wheel and variations between the filters.

Anatomy of a Filter Cube-Emission, Dichroic, ExcitationWithout a multi-view attachment or emission filter wheel, researchers must fall back on a third hardware configuration for FRET, which involves using a separate filter cube for both donor and acceptor fluorescence. In this configuration, while each cube has an exciter, dichroic and emitter, it is extremely important to remember that the exciter and dichroic in both sets are identical. While this hardware configuration allows for the discreet collection of donor and acceptor fluorescence, it is slow (usually dependent on users explicitly changing filter cubes) and the most susceptible to subtle image registration shifts. It is most appropriate for fixed samples or those that are changing very slowly over time.

Grating based fluorescence spectroscopy is another useful method to perform FRET. For FRET to occur, there must be significant overlap between the emission spectrum of the donor and the excitation spectrum of the acceptor. A fluorescence spectrometer can accurately characterize this spectral overlap, ensuring that the chosen fluorophore pair is suitable for FRET experiments. This analysis helps in selecting the optimal fluorophore pairs for specific applications. Advanced fluorescence spectrometers equipped with time-resolved capabilities can measure fluorescence lifetimes. By comparing the donor's fluorescence lifetime in the presence and absence of the acceptor, researchers can calculate FRET efficiency more accurately. A decrease in the donor's lifetime in the presence of the acceptor confirms energy transfer.

Holographic Diffraction Grating Production.
Explore the Features

Full Optical Capabilities

Learn more about the full suite of capabilities which are trusted by leading Global OEM’s and their Start-up challengers to deliver the right photons, to the right place, at the right time.

Learn More

Omega Optical Forster Resonance Energy Transfer FRET

Key Parameters

Bandpass center wavelength

Peak Transmission is the maximum transmission of the filter (%).
Center Wavelength (CWL) is the midpoint of the transmitting region (nm).
Full-width half-max (FWHM) describes the width of the transmitting region at half of the maximum transmission value (nm). Because of the physics of thin-film interference filters, FWHM generally increases with wavelength. It can be calculated using a percentage of the center wavelength.

Tolerancing

Cut-on and Cut-off wavelengths can be used instead of FWHM and CWL to designate the edges of the filter's transmitting region (nm).
Pass band ripple describes the flatness of the transmitting region (%). Some narrow bandpass filters and older designs have a peaked pass band shape instead of a flat top. This depends on filter design. Typically flat-top designs require more layers and greater cost.

Blocking

Blocking specifications determine the wavelength range (nm) and blocking (OD) of unwanted light. Typically the wavelength range is dictated by the response curve of your detector. However, extended blocking is not required for some applications (for example laser-cleanup or Raman filters) where unwanted signals are only present in a limited wavelength range. Optical density (OD) is a log scale relating the transmitted to incident light.
OD=2-log(%T)

Edge Refinement

Edge Steepness describes the transition from blocking to transmission in more detail (nm).

Key supporting capabilities

Solution

Our commitment to finding the right solutions for difficult problems means that we are capable of doing what other engineers cannot. Some say we bend physics and solve the unsolvable. Can we? Maybe. What we know we can do is this: We ask questions and we dig, because sometimes the solutions people believe they need only scratch the surface.

Solution

Omega coating heritage dates back to 1936, and spans a wide range of technologies: Sunglasses to space-borne sun-visors; halogen light bulbs to COVID-fighting UV-C lamps, supermarket scanners to modern printers, astronomical mirrors to prestigious art installations to entertainment optics: if an application requires a precision or decorative coatings we surely have a proven solution.

Solution

We believe you cannot confidently make what you cannot measure, and therefore have invested in standard and custom metrology and inspection stations. Certified ISO 9001:2015. Lean manufacturing, including 6 Sigma methodologies and approaches to problem solving.

Solution

We keep one eye on the horizon for the needs of tomorrow. We are Trusted Collaborators with a passion for innovation.

Related Applications

Application
FTIR
Application
Flourescence
Application
PCR
Application
Raman/ Sensing and Illumination

Latest news

Latest
Collaborate with US

Request a Quote

Talk with US

Have a Question

Work with Us

Join
Our Team